Product information
Maintenance
After 2,000 hours of operation, or once a year, The device Kinetec Centura B&W™ requires lubrication and maintenance operations
(lubrication of the joints and pointer stops). The need for maintenance is indicated by display of the message SERVICE TIME Mx when the
system is switched on.
Despite this indication, you can continue to use your machine by pressing [START], but you should contact your nearest KINETEC
to have the maintenance operations carried out as soon as possible.
An after-sales service inspection sheet and the technical catalogue are available on request from your KINETEC
:
WARNING
:
WARNING
Before using this machine, always check that the machine is not damaged, in particular the protective housings.
When the machine is no longer in working condition, please return it to us, together with its accessories, for destruction.
Trouble shooting
A spare parts list and technical catalogue are available on request from your KINETEC
If, after connecting the power supply cable to the power supply and
switching on the machine:
• The display does not indicate any information:
- Check that the electrical socket is live using another device or voltmeter.
- Replace the fuse(s) of the connector with fuses of the same type and calibre:
2 fuses T 750 mA 250V (6.3 x 32) (KINETEC
- If the display still does not indicate any information, contact your nearest KINETEC
technician.
• Your KINETEC
does not work and the display indicates "50 STOP 25 115", press START again.
®
If your machine still does not work, contact your nearest KINETEC
• Your machine does not work and the display indicates:
"SERVICE D1": angle measurement function failure,
or "SERVICE D2": no movement,
or "SERVICE D3": abnormal electrical consumption by motor,
or "SERVICE D4": power failure or disconnected motor,
Contact your nearest KINETEC
Cleaning
Before carrying out any cleaning operation, SWITCH OFF the unit and disconnect the power supply. In order to ensure optimal hygiene, you
are advised to clean the machine for each new patient.
Cleaning should be carried out in the environmental conditions specified in the "Technical Specifications" section below.
Use a DISINFECTANT product (alcohol-free or <5% alcohol solution) in spray.
FOR OPTIMAL HYGIENE, A NEW SET OF PADS SHOULD BE USED FOR EACH PATIENT.
All the consumables enable hazard-free disposal.
Disposal and recycling
a • Packaging: The packaging must be separated into plastic and paper / cardboard components and taken to special recycling sites.
b • KINETEC
patient pad kit: Clean with a disinfectant product then take it to special recycling sites.
®
c • Unit: It contains electronic components, cables, aluminium, steel and plastic parts. When the machine is no longer operational,
disassemble it, separate it into different types of material and take these to authorised recycling centres or return the machine to Kinetec
SAS for destruction. Or contact the local authorities to determine the appropriate method of disposal for parts and accessories that are
potentially hazardous to the environment.
Kinetec Centura B&W™
Before using this machine, always check that the electrical socket is in good condition and is suitable for the
splint power supply cord. Only use the original cable supplied with the machine. Check that the cables remain
free around the device so that they do not get damaged.
order: 4610007434).
®
technician if the same message is displayed after having witched the device off, then on.
®
distributor.
®
®
technician.
®
21/24
technician
®
distributor.
®
17