SYMBOLS/
SYMPTOMS
No power-on by
AVG
pressing
button
AVG
Measuring Error
Symbol shown
AVG
AVG
on display
or the blood
pressure value
is displayed
excessively low
or high.
Measuring
Error Symbol
Measuring
Error Symbol
Measuring
Error Symbol
Note: If "EP" appears on the display, just return the device to your local
distributor or importer.
CONDITIONS/
CAUSES
Worn-out batteries.
Battery polarities
have been positioned
incorrectly.
Cuff has been placed
incorrectly.
Did you talk or move
during measurement?
Shaking of the wrist
with the cuff on.
Air circuit adnormality.
Cuff tube may not be
plugged into monitor
correctly.
Inflation pressure
exceeding 300 mmHg.
Can't determine blood
pressure measurement
data.
14
INDICATION/
CORRECTION
Replace them with
2 new AAA (LR03)
alkaline batteries.
Re-insert the batteries
in the correct
positions.
Wrap the cuff properly
so that it is positioned
correctly.
Measure again. Keep
wrist steady during
measurement.
Check cuff connection.
Measure again.
Switch the unit off,
then measure again.
Wrap the cuff properly
and keep steady.
Measure again.