Changing The Water; Cleaning The Spray Tubes; Labormat Sd Accessories - Dreve LABORMAT SD Instrucción

Ocultar thumbs Ver también para LABORMAT SD:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

Labormat SD

7.3 Changing the water

In order to obtain consistently good
results when cleaning and boiling out
models, the wax dissolved in water
should be removed regularly (once a
week, more often if used more fre-
quently). Do this by setting the tempera-
ture controller (24) to 0 °C or switching
off the device and allowing it to cool
(ideally overnight).
The wax dissolved in the water auto-
matically collects as a layer on the sur-
face. After removing the flask baskets
and support plates (55), this layer can
be scooped out along with any other
impurities floating on the surface.
When giving the device an overall and
thorough cleaning (min. once a week,
more frequently if used more often), the
device can be switched back on again
after scooping out the wax layer and
preheated to 65 °C.
The water inside the device can be
pumped off using the discharge hose
supplied. Do this by removing a spray
tube (47) (see next section) and estab-
lishing the tube/Labormat SD con-
nection with the aid of the bayonet
coupling attached to the tube. Place
the other end of the tube in a drain
(e. g. sink).
Activate the pump with the aid of the
timer (18) or hold button (15). The
water is automatically pumped off until
3–4 cm sediment remains on the base.
The heavily-soiled residual water is not
pumped off but rather discharged via a
drainage valve (42) located on the base
50
of the water tank. Fully open the lower
front cover (5) (magnetic catch) and
drainage valve (42) using the red shut-off
valve. The residual water flows off and
is collected in the collection basin (54)
underneath. Remove rough deposits
and deposits adhering to the inside
of the water tank and rinse the device
with a little water. Close the drainage
valve (42). Detach the collection basin
(54) from the device and dispose of the
waste water. Fill with fresh water as far
as the maximum mark outlined in the
section on „Filling with water".
Important: Only dissolve Unisol W
in cold water; high water tempera-
tures cause foaming.

7.4 Cleaning the spray tubes

Pump mode can also enable rough impu-
into the spray tube (47) and use rotary
rities to penetrate the spray tubes (47)
and thrusting movements to release and
along with the water. This can impair
remove rough dirt and calcium deposits.
the function of the spray nozzles and
The tube should then be rinsed under
should be checked at regular intervals,
running water.
cleaning if necessary.
The spray tubes are reassembled in the
This necessitates removal of the spray
reverse order to the process described
tubes (47) which are connected to
above.
the water line in the device via bayo-
net catches. Unlock the bayonet catch
Important: When inserting the
by a half-turn to the right and gently
spray tubes, ensure that the spray
pull the entire spray tube (47) out at
nozzles face downwards.
an upward incline. A cleaning brush is
supplied with each new device. Clean
by releasing the dummy plug (48) at
the end of the tube, inserting the brush

8. Labormat SD accessories

Waste water
Fresh water
discharge hose
supply hose
Item no.: D51142KPL
Item no.: D1017E
Cleaning brush for
spray tubes
Item no.: 51074
51