12. Cleaning and maintenance

12.1 Checking the water level

Labormat devices are hot water devices
which release steam during the spray
process and when the lid is open. This
lowers the water level and necessi-
tates checking the water level on a
daily basis.
If the water level reaches the level
of the heating cover (6), it is im-

12.2 Restoring operational readiness

Important: Always pull the mains
plug before opening the device.
The overheating protection safety
switch (21) is located on the control
panel (7) and is secured with a pro-
tective cap.
If reactivation is necessary, release the

12.3 Changing the water

In order to obtain consistently good
results when cleaning and boiling out
models, the wax dissolved in water
should be removed daily.
Do this by setting the temperature con-
troller (19/22) to 0 °C or switching off
the device and allowing it to cool (ide-
ally overnight).
The wax dissolved in the water auto-
matically collects as a layer on the sur-
face as the water cools down. After
removing the flask baskets and support
plate (55), this layer can be scooped out
along with any other floating impuri-
58
perative that water is added
(Please refer to the section „Filling
with water"). If the water level falls to
the level of the heating element (28),
a safety switch (21) isolates the entire
device from the mains supply to protect
against overheating.
protective cap in an anticlockwise direc-
tion and press the green locking but-
ton underneath until it audibly clicks
into place.
Immediately replace the protective cap
on the locking button on account of
the risk of electrical components being
damaged by water.
ties. Give the device an overall and thor-
ough cleaning at least once a week and
more frequently if used more often after
scooping out the wax layer. Discharge
the residual water through the drain-
age valve (43) on the bottom left of
the rear wall. The valve (43) is opened
by turning the red regulating wheel
to the left on the left housing wall.
The waste water can be collected in
a bucket or directed into the sewerage
system through a hose (only possible
with the corresponding incline). Close
the drainage valve (43) and remove any
rough impurities from the water tank.
solubility and avoiding emulsification of
Fill with fresh water as far as the maxi-
the water with residual wax.
mum mark outlined in the section on
„Filling with water".
Important: Only dissolve Unisol W
in cold water; high water tempera-
We recommend adding Unisol W
(see „Accessories") for improved wax
tures cause foaming.

12.4 Cleaning the spray tube

Pump mode can also enable rough impu-
Clean by removing the dummy plug (48)
rities to penetrate the spray tube (47)
at the end of the tube, inserting the
along with the water. This can impair
brush into the spray tube and use rotary
the function of the spray nozzles and
and thrusting movements to release and
should be checked at regular intervals,
remove rough dirt and calcium deposits.
cleaning if necessary.
The tube should then be rinsed under
This necessitates removal of the spray
running water.
tube (47) which is connected to the
The spray tube is reassembled in the
water line in the device via a bayonet
reverse order to the process described
catch (45). Unlock the bayonet catch by
above.
a half-turn backwards and gently pull
the entire spray tube out at a forward/
Important: When inserting the
upward incline. A cleaning brush is sup-
spray tube, ensure that the spray
plied with each new device.
nozzles face downwards.
59