WIKA TR33 Manual De Instrucciones página 30

Ocultar thumbs Ver también para TR33:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

12. Specifications
Output signal (4 ... 20 mA version)
Factory configuration
Measuring range
EN
Current signals for
error signalling
Current value for
sensor short-circuit
Communication
Info data
Configuration and
calibration data
Configuration software WIKAsoft-TT
Voltage supply
Supply voltage U
Supply voltage input
Permissible residual
ripple of supply voltage
Time response
Switch-on delay,
electrical
Warm-up time
30
Measuring range 0 ... 150 °C [32 ... 302 °F]
Other measuring ranges are adjustable
Configurable in accordance with NAMUR NE 043
downscale ≤ 3.6 mA
upscale ≥ 21.0 mA
Not configurable in accordance with NAMUR NE 043
downscale ≤ 3.6 mA
Tag no., description and user message can be stored
in transmitter
Permanently stored
→ Configuration software (multilingual) as a
download from www.wika.com
DC 10 ... 30 V
B
Protected against reverse polarity
10 % generated by U
current
Max. 4 s (time before the first measured value)
After approx. 4 minutes, the instrument will function
to the specifications (accuracy) given in the data
sheet.
WIKA operating instructions model TR33
< 3 % ripple of the output
B

Publicidad

Tabla de contenido
loading

Tabla de contenido