12. Specifications
Operating conditions
Ambient temperature range
Version 4 ... 20 mA (TR33-Z-TT)
Version Pt100 (TR33-Z-Px) / Pt1000
(TR33-Z-Sx)
Storage temperature range
Climate class per IEC 60654-1
Version 4 ... 20 mA (TR33-Z-TT)
Version Pt100 (TR33-Z-Px) / Pt1000
(TR33-Z-Sx)
Maximum permissible humidity,
condensation
Maximum operating pressure
For protection tube Ø 3 mm [0.12 in]
For protection tube Ø 6 mm [0.24 in]
Salt fog
Vibration resistance per IEC 60751
Shock resistance per IEC 60068-2-27
Maximum permissible autoclaving
conditions
WIKA operating instructions model TR33
-40 ... +85 °C [-40 ... +185 °F]
Version with FKM O-ring: -20 °C [-4 °F]
-50 ... +85 °C [-58 ... +185 °F]
Version with FKM O-ring: -20 °C [-4 °F]
-40 ... +85 °C [-40 ... +185 °F]
Version with FKM O-ring: -20 °C [-4 °F]
Cx (-40 ... +85 °C [-40 ... +185 °F],
5 ... 95 % r. h.)
Version with FKM O-ring: -20 °C [-4 °F]
Cx (-50 ... +85 °C [-58 ... +185 °F],
5 ... 95 % r. h.)
Version with FKM O-ring: -20 °C [-4 °F]
100 % r. h., condensation allowed
1) 2)
140 bar [2,030 psi]
270 bar [3,916 psi]
IEC 60068-2-11
10 ... 2,000 Hz, 20 g
50 g, 6 ms, 3 axes, 3 directions, three
times per direction
Max. 134 °C, 3 bar abs., 100 % r. h.,
duration 20 min., max. 50 cycles
Autoclavable with mounted protective
cap at coupler connector
1)
31
EN