If the two spots are at different heights, mark a spot half way between
the two (c). Remove the two calibration hole covers by using a screw-
driver. Insert a hex screw (1,5 mm) into the side calibration hole (X) and
turn the adjusting screw until the laser dot is on the new mark you
made (c).
4.
Turn the SP5 so that the front beam (2) is pointed at the mark you made.
If the laser dot is not lined up with the mark, insert the adjusting wrench
in the front calibration hole (Y) and turn the adjusting screw until the
laser dot is on the mark you made.
5.
Start at step 1 again and re-check your adjustments. If the three laser
dots stay on the same mark, then calibration is complete.
6.
Push on the calibration port covers.
SAFETY AND CERTIFICATIONS
Working safely with this instrument is possible only when the operating and
safety information are read completely and the instructions contained there-
in are strictly followed.
Do not remove the label on the side of the housing.
The use in combination with other optical instruments, manipulations or use
in other applications other than described here, can lead to dangerous laser
outputs.
Do not stare into the laser beam.
Do not direct the laser beam at other persons. Since the laser beam is of the
boundled type, check the beam path over a relatively long distance and take
the necessary precautions.
This laser complies with all applicable portions of title 21 of the Code of
Federal Regulations set by the Dept. of Health, Education, and
Welfare, the Food and Drug Administration, the Center for Devices, and the
Bureau of Radiological Health.
8 • SP5
Your SP5 is calibrated and ready to go.