Nebuliser
Preparation
• Pull the two tubes and the adapter off the nebuliser.
• Make sure that all residual volume is removed from the
nebuliser.
• Detach the adapter from the tube.
• Dismantle the nebuliser into its individual parts.
Cleaning
The PARI SINUS tubing system cannot be cleaned or disin-
fected. Regarding care of the tubing system, follow the in-
structions in the corresponding section [see: Care of the
PARI SINUS tubing system, page 31].
• Briefly rinse all parts used in running drinking water before-
hand.
• Place all disassembled com-
ponents in warm drinking water
with a little dishwashing liquid
for about 5 min.
• Rinse off all parts thoroughly in drinking water.
• Shake the water out of all of the parts.
Disinfecting
After cleaning, disinfect all of the disassembled parts (only
parts that have been cleaned can be disinfected effectively).
The recommended disinfection procedures are described be-
low. Descriptions of other validated disinfection procedures are
available from PARI GmbH upon request.
The PARI SINUS tubing system cannot be cleaned or disin-
fected. Regarding care of the tubing system, follow the in-
structions in the corresponding section [see: Care of the
PARI SINUS tubing system, page 31].
– 27 –