Hand Held Remote Installation
and Operation
The Hand Held Remote Control allows the
operator full functionality of the 10k Pump from
within the sterile field.
WARNINGS:
1.
Prior to installing the remote, ensure the
contact surfaces of the connector are dry.
Moisture in the connector can cause
system malfunctions.
2.
Prior to using the remote, ensure it is
cool to the touch. A hot remote can cause
system malfunctions and lead to patient
injury.
NOTE:
The remote may also be bagged
during use to keep the device sterile.
1.
Use surgical aseptic technique to connect
the remote to the pump.
(a) Circulator - Hold the sterilization
container in the vicinity of the Scrub.
(b) Scrub - Remove the remote from the
sterilization container and place it in
the sterile field.
(c) Scrub - Hand the connector body to
the Circulator. The remainder of the
remote and cord stays in the sterile
field.
(d) Circulator - Line up the alignment
marks on the connector body and
remote receptacle on the pump, and
insert the connector completely.
Flow Buttons — Press either of these
buttons to toggle between the preset
High (2000 ml/min. - "HI" will display)
and Low (1000 ml/min. - "LO" will
display) flow rates. Note: This button has no
functionality when the Model 10k200
Laparoscopic Tubing Set is in use.
Tamponade Assist Button — Press for a
20 mmHg increase in set pressure. After
two minutes, set pressure automatically
returns to the original set point. Each
consecutive press of the button results in another
20 mmHG increase. Pressing any other button
returns the pump to the original set point. Note:
This button has no functionality when the
Model 10k200 Laparoscopic Tubing Set is in
use.
Pressure Decrease Button — Press to
decrease set pressure.
Pressure Increase Button — Press to
increase set pressure.
"Run/Stop" Button — Press to start
the pump roller when the Standby
switch is in the on position. Press
again to stop the pump roller.
E-9
E