Remote Cleaning

1.
Follow universal precautions for protective
apparel when handling and cleaning
contaminated equipment.
2.
Prior to any cleaning, sterilization or
disinfection, the door must be opened to a
partially opened position.
3.
To clean, immerse remote in warm, soapy
water and scrub with a soft brush.
Method
Steam (wrapped)
Steam (wrapped)
Steam (wrapped)
All Cycles: 8 Minute Dry Time
3.1.3
WARNING: The use of disinfecting
solutions for an exterior instrument wipe will
not sterilize the remote and is not
recommended.
NOTES:
1. Guidelines provided below do not
2. Additional air drying time may be
3. Sterilization validation is based on
Cycle
Minimum Temperature
Pre-vacuum
Gravity
Gravity
E-15

Remote Sterilization

guarantee the device is sterile after the
procedure. Your institution is still
responsible for the normal sterility
assurance validation.
required for complete heat and moisture
dissipation. Operation of a remote that is
not completely cool and dry may
decrease its performance and or
reliability.
AAMI guidelines (Association for the
Advancement of Medical
Instrumentation).
Minimum Exposure
270°F (132°C)
270°F (132°C)
250°F (121°C)
E
4 minutes
8 minutes
30 minutes