(c) To continuously adjust the irrigation
pressure, press and hold the
E
appropriate button. The pressure will
increase in increments of 5 for three (3)
seconds, then accelerate to increments
of 15.
(d) To quickly adjust the pressure, press
and release the appropriate button. The
pressure increases or decreases as fast
as the button is pressed and released, in
increments of 5.
9. To begin fluid flow, open the inflow clamp.
NOTE: if using the 10k200 Tubing Set,
press the blue inflow button.
10. To stop the pump, press the "Run/
Stop" button again, either on the
pump or the remote control.
11. When the procedure is complete,
remove the Tubing Set by pulling
down on the cassette and removing
the tubing loop from around the pump rotor.
Dispose of Tubing Set properly.
2.5

Programming the 10k Console

The 10k Console is capable of two display
modes: "Set point" and "Pressure feedback".
The console is shipped from the factory in the
"Set point" mode.
1. If the "Pressure feedback" mode is
preferred, first ensure that there is not a
tubing set loaded. Ensure the power is off
and perform the following steps:
(a) Press and hold the Pressure
Down, Pressure Up, and Run/
Stop buttons simultaneously
while turning the console
power on. Release all buttons
once the display indicates
"d 1".
(b) Place a finger on the face of
the pump to the right of the standby
power switch, and to the left of the
pressure sensor window. While
applying slight pressure, slide your
finger toward the pressure sensor
window until the display indicates
"Pr".
E-12