5. Set-up

+
The sensION
MM 150
- Temperature coefficient: 2.00%/°C.
- Reference temperature: 25°C.
- TDS calculation factor: 0.64.
To display and/or modify these parameters, follow the diagram below.
5.1. Temperature
coefficient (TC)
Enter the relevant coefficient value (0.00
to 9.99%/°C).
Recommended value for aqueous sam-
ples: 2.00%/°C.
5.2. Reference
temperature (RT)
Select a value between 20 and 25°C.
5.3. TDS factor (F)
Enter the relevant value
(0.00 to 4.44).
10
sensION+ MM150 HACH 0710.indd ANG:10
sensION+ MM150 HACH 0710.indd ANG:10
is supplied configured with the following parameters: