View instrument information

1. Select INSTRUMENT INFORMATION.
2. Select an option.
Option
ANALYZER INFO
MODULE INFO (option is
available when a module is
installed)

Set up LINK2SC

The LINK2SC procedure is a secure method for the data exchange
between process probes and LINK2SC-compatible photometers. Use an
SD memory card for the data exchange. Refer to the LINK2SC
documentation on
http://www.hach.com
LINK2SC procedure.
1. Select LINK2SC.
2. Select an option.
Option
Description
CREATE A NEW
Starts the grab sample operation for the measurement
JOB
value exchange between the analyzer and the
laboratory.
JOB LIST
Selects the job file to send the job to the laboratory or
deletes the job. JOB TO LAB—the analyzer data is sent
to the SD card as a job file. ERASE JOB—deletes the
data.
JOB ID MIN
Specifies the minimum value for the ID number range.
JOB ID MAX
Specifies the maximum value for the ID number range.
18 English
Description
Shows the software information and
serial number.
Shows the connected modules with
software information and serial number.
for a detailed description of the
Use an SD card
Use an SD memory card to update the software and firmware and to
download the event and data logs. The SD icon is visible in the upper
status bar of the main measurement screen when a card is installed. The
manufacturer recommends to use an SD card with a minimum of 2 GB
storage capacity.
1. Install the SD card (refer to
2. Select SD CARD SETUP from the MAIN MENU.
Note: The SD CARD SETUP option shows only when an SD card is installed.
3. Select an option.
Option
Description
UPGRADE
Shows when an the upgrade file is available on the
SOFTWARE
SD card. Select the specific device for the
(conditional)
upgrade.
SAVE LOGS
Selects the device for the download data and to
save the logs for last day, last week, last month or
all.
MANAGE
Saves and restores the backup settings, restore
CONFIGURATION
settings and/or transfer settings between
instruments.
WORK WITH DEVICES READ DEVICE FILES—selects the data for each
device to save on the SD card. Options: sensor
diag, measurement data (curve data for a
measurement cycle), cal history, cal data and/or
test script. WRITE DEVICE FILES—shows when
an upgrade file is available for a new measurement
cycle script.

Update the firmware

Use an SD card with an upgrade file to update the firmware for the
controller, sensor or network card. The upgrade menu is shown only
when the SD card contains an upgrade file.
1. Install the SD card into the SD card slot.
2. Select SD CARD SETUP from the MAIN MENU.