Medtronic ITREL 3 Manual De Implante página 9

Neuroestimulador
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 45
7425IPG_ch.fm 4/27/04 2:22 pm
UC200xxxxxx EN
4.625 x 6 inches (117 mm x 152 mm)
Table 1. Shipping, operating, and power-on-reset (POR) values for the
Model 7425 Itrel 3 Neurostimulator
Programmable parameter
Polarity
Program
a
All measurements are approximate.
b
Power-on-reset (POR) turns OFF stimulation by resetting the amplitude to 0.0 V and all
electrodes to OFF. POR occurs when there is a temporary fluctuation in battery voltage (eg,
due to electromagnetic interference during electrocautery or defibrillation) or the battery is
depleted. When POR occurs, the serial number is displayed as "????" on the clinician
programmer.
c
Exposure to a strong source of electromagnetic interference (EMI) can cause the output
switch to toggle On or Off. However, since the amplitude setting is 0.0 V, there is no
neurostimulator output.
221656002
Rev A
Shipping
Operating
Electrode
number
0
OFF, –, or +
1
OFF
OFF, –, or +
2
OFF
OFF, –, or +
3
+
OFF, –, or +
Case
OFF
OFF, or +
Medtronic Confidential
NeuroImplant_R01
a
. (continued)
b
POR
OFF
OFF
+
OFF
7425 2004-05
English 7

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

7425

Tabla de contenido