Table 1. Shipping, operating, and power-on-reset (POR) values for the
Model 7425 Itrel 3 Neurostimulator
Programmable parameter
Polarity
Program
a
All measurements are approximate.
b
Power-on-reset (POR) turns OFF stimulation by resetting the amplitude to 0.0 V and all
electrodes to OFF. POR occurs when there is a temporary fluctuation in battery voltage (eg,
due to electromagnetic interference during electrocautery or defibrillation) or the battery is
depleted. When POR occurs, the serial number is displayed as "????" on the clinician
programmer.
c
Exposure to a strong source of electromagnetic interference (EMI) can cause the output
switch to toggle On or Off. However, since the amplitude setting is 0.0 V, there is no
neurostimulator output.
221656002
Rev A
Shipping
Operating
Electrode
number
0
–
OFF, –, or +
1
OFF
OFF, –, or +
2
OFF
OFF, –, or +
3
+
OFF, –, or +
Case
OFF
OFF, or +
Medtronic Confidential
NeuroImplant_R01
a
. (continued)
b
POR
–
OFF
OFF
+
OFF
7425 2004-05
English 7