WARNING:
When doing the control solution test, you have to mark it so that the test result
will NOT mix with the blood glucose TEST RESULTS stored in the memory. Failure
to do so will mix up the blood glucose test results with the control solution test
results in memory.
3. Apply control solution
Shake the FORA control solution vial thoroughly before use. Squeeze
out first drop and wipe it off, then squeeze out another drop and place
it on the tip of the vial cap. Hold the meter to move the absorbent hole
of the test strip to touch the drop. Once the confirmation window fills
completely, the meter will begin counting down. To avoid contaminating
the control solution, do not directly apply control solution onto a strip.
4. Read and compare the result
After counting down to 0, the control
solution test result will appear on the
display. Compare this result with the
range printed on the test strip vial and it
should fall within this range. If not, please
Normal
read the instructions again and repeat
the control solution test.
118-160