Out-of-range results
If you continue to have test results fall outside the range printed on the
test strip vial, the meter and strips may not be working properly. Do NOT
test your blood. Contact the local customer service or place of purchase
for help.
NOTE:
• The control solution range printed on the test strip vial is for control solution
use only. It is not a recommended range for your blood glucose level.
• See the MAINTENANCE section for important information about your control
solutions.
• It is recommended that the control solution test should be done at room
temperature (68 ° F to 77 ° F/20 ° C to 25 ° C). Make sure your control solution,
meter, and test strips are at this specified temperature range before testing.

TESTING WITH BLOOD SAMPLE

WARNING:
To reduce the chance of infection:
• Never share a lancet or the lancing device.
• Always use a new, sterile lancet. Lancets are for single use only.
• Avoid getting hand lotion, oils, dirt, or debris in or on the lancets or the
lancing device.
• Wash and dry your hands thoroughly before and after handling the meter,
lancing device and test strips to prevent infection. For more information,
please refer to the Cleaning and Disinfection section.
• If the meter is being operated by a second person who is providing testing
assistance to the user, the meter and lancing device should be
decontaminated prior to use by the second person.
Sharing the lancing device and lancets may increase the risk of
spreading infectious diseases. Lancing device must not be used for
more than one person.