Hach PHOSPHAX sigma Manual Básico Del Usuario página 102

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 113
SETTINGS
Setting of all device parameters: This includes measurement range, correction factor, current output, date, time etc.
LANGUAGE (DE, EN, FR, NL)
STANDARD (concentration of the standard solution used: 1.95–2.05 in 0.01 increments)
SPECIAL P (Total P)
SPECIAL OP (Ortho-P)
P/OP (number of measurement cycles per parameter each 0–10)
RELAY-MIN (lower limit-value contact 0.00 to [CURRENT RANGE P] setting)
RELAY-MAX (upper limit-value contact 0.00 to [CURRENT RANGE P] setting)
SIGMATAX (sample feed via SIGMATAX: yes, no)
DATE (setting of the date)
TIME (setting of the time)
CLEANING (automatic cleaning off, 6 h, 12 h, 1–7D, start time for each hourly --> last Display of last cleaning)
CALIBRATION (automatic calibration off, 1–7D, start time for each hourly --> last Display of last calibration)
VERSION (program version)
PASSWORD (password protection on activation 4-digits using digits 1-4)
SERVICE STATION (place and contact partner)
BACKLIGHT (display lighting)
English 102
FACTOR (correction factor for the total phosphorus measurement values
0.20–5.00 in 0.01 increments)
CURRENT RANGE (measurement range for current output 1
0.5–5.0 in 0.1 increments)
CURRENT (current transfer for current output 1 0–20 mA, 4–20 mA)
MALFUNCTION (current transfer in the event of a fault with current output 1 off
(= last measurement value), 0 mA, 20 mA)
FACTOR (correction factor for the orthophosphate measurement values
0.05–1.0–20.0)
CURRENT RANGE (measurement range for current output 2
0.5–5.0 in 0.1 increments)
CURRENT (current transfer for current output 2 0–20 mA, 4–20 mA)
MALFUNCTION (current transfer in the event of a fault with current output 2 off
(= last measurement value), 0 mA, 20 mA)

Publicidad

Tabla de contenido
loading

Tabla de contenido