Accuracy Table

3.3 Accuracy Table
(1) AUTO A (AC/DC current automatic detection)
During AC detection: Conforms to accuracy specifications described in "(4) AC+DC current" (p. 44).
During DC detection: Conforms to accuracy specifications described in "(3) DC current" (p. 43).
(2) AC current
Measurement value/MAX/MIN/AVE
Accuracy guarantee range
(Resolution)
1.0 A to 30.0 A (0.1 A)
30.1 A to 900.0 A (0.1 A)
900.1 A to 999.9 A (0.1 A)
42
Accuracy guarantee
frequency range
10 Hz ≤ f < 45 Hz
45 Hz ≤ f ≤ 66 Hz
66 Hz < f ≤ 1 kHz
10 Hz ≤ f < 45 Hz
45 Hz ≤ f ≤ 66 Hz
66 Hz < f ≤ 1 kHz
10 Hz ≤ f < 45 Hz
45 Hz ≤ f ≤ 66 Hz
66 Hz < f ≤ 1 kHz
Measurement accuracy
Filter OFF
±1.8% rdg.±1.0 A
±1.3% rdg.±0.8 A
±2.0% rdg.±1.0 A
±1.8% rdg.±0.5 A
±1.3% rdg.±0.3 A
±2.0% rdg.±0.5 A
±2.3% rdg.±0.5 A
±1.8% rdg.±0.3 A
±2.5% rdg.±0.5 A
Filter ON
±2.3% rdg.±1.0 A
±1.8% rdg.±0.8 A
±2.3% rdg.±0.5 A
±1.8% rdg.±0.3 A
±2.8% rdg.±0.5 A
±2.3% rdg.±0.3 A