Accuracy Table
(4) AC+DC current
Measurement value/MAX/MIN/AVE
Accuracy guarantee
range (Resolution)
1.0 A to 30.0 A (0.1 A)
30.1 A to 900.0 A (0.1 A)
900.1 A to 999.9 A (0.1 A)
PEAK MAX/PEAK MIN
Accuracy guarantee range (Resolution)
±10 A to ±1000 A (1 A)
±1001 A to ±1500 A (1 A)
44
Accuracy guarantee
frequency range
10 Hz ≤ f < 45 Hz
DC, 45 Hz ≤ f ≤ 66 Hz
66 Hz < f ≤ 1 kHz
10 Hz ≤ f < 45 Hz
DC, 45 Hz ≤ f ≤ 66 Hz
66 Hz < f ≤ 1 kHz
10 Hz ≤ f < 45 Hz
DC, 45 Hz ≤ f ≤ 66 Hz
66 Hz < f ≤ 1 kHz
Accuracy guarantee
frequency range
10 Hz ≤ f < 45 Hz
DC, 45 Hz ≤ f ≤ 66 Hz
66 Hz < f ≤ 1 kHz
10 Hz ≤ f < 45 Hz
DC, 45 Hz ≤ f ≤ 66 Hz
66 Hz < f ≤ 1 kHz
Measurement accuracy
Filter OFF
±1.8% rdg.±1.2 A
±1.3% rdg.±1.8 A
±2.0% rdg.±1.2 A
±1.8% rdg.±0.7 A
±1.3% rdg.±1.3 A
±2.0% rdg.±0.7 A
±2.3% rdg.±0.7 A
±1.8% rdg.±1.3 A
±2.5% rdg.±0.7 A
Measurement accuracy
±1.8% rdg.±7 A
±1.3% rdg.±7 A
±2.0% rdg.±7 A
±2.3% rdg.±7 A
±1.8% rdg.±7 A
±2.5% rdg.±7 A
Filter ON
±2.3% rdg.±1.2 A
±1.8% rdg.±1.8 A
–
±2.3% rdg.±0.7 A
±1.8% rdg.±1.3 A
–
±2.8% rdg.±0.7 A
±2.3% rdg.±1.3 A
–