3.1. General rules
E120
E123
E128
E129
E130
E131
E132
E133
E134
E135
E136
E137
62
Overrange of the primary measured value (pH, U, Ipol, Upol or T
with MEAS T). The secondary measured value (temperature) may
be instable as well.
Exit: Correct error or &m $S.
Missing EP for calculation.
Exit: The error message disappears on next startup or on
recalculation.
No new mean.
Exit: The error message disappears on next startup or on
recalculation.
No new common variable, old value remains.
Exit: The error message disappears on next startup or on
recalculation.
Wrong sample. For SET with preset titration direction the first
measured value lies behind the endpoint.
Exit: The error message disappears on next startup.
No EP set for SET.
Exit: The error message disappears on next startup.
Silo empty and it has been started with open silo or empty silo has
been opened.
Exit: Send a silo entry.
Silo full.
Exit: Send new command.
No method. A method, which is required from the silo memory, does not
exist.
Exit: The error message disappears on next startup.
Check temp.sensor in MEAS T.
Exit: Correct error or &m $S.
Same buffer in CAL. Measured value of the second buffer differs less than 6
mV from the measured value of the first buffer.
Exit: Correct error or &m $S.
XXX Bytes are missing so that the method, the silo line could not be stored.
Exit: Send new command.
719 S Titrino