Riester ri-former RPT-100 Instrucciones Para El Uso página 26

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

Typical measurement times
(after insertion into measurement site):
Oral (standard mode):
Axillary mode:
Rectal mode:
Direct mode (all sites):
Typical times for switch to direct mode:
If no measurement site detected:
If temperature does not stabilize:
Water bath accuracy (35.5°C to 42.0°C):
Direct mode (all sites):
Standard prediction mode (all sites)**:
Quick prediction mode (oral only)**:
**Greater than 95% of prediction mode
readings are within specified accuracy
Patient accuracy:
Probe:
Isolation chamber:
Device materials:
ALL MATERIALS ARE LATEX-FREE.
Type of protection against electrical shock:
Degree of protection against electrical shock:
Mode of operation:
Degree of safety in the presence of:
Transport and storage:
Environmental conditions for use:
Temperature:
Barometric pressure:
IP-Index:
Input:
Output:
The RPT-100 Predictive Thermometer extension module is in compliance with
the following international regulatory and safety standards:
ISO 60601-1: 2005
IEC 60601-1-2: 2014
IEC 60601-1-6: 2010
IEC 62366: 2007
ASTM E 1112-00: 2011
ASTM E 1104-98: 2009
ISO 80601-2-56
ISO 10993-5: 2009
ISO 10933-10: 2010
26
Oral (quick mode): 3-5 seconds
(non-fever temps), 8-10 seconds
(fever temps)
6-10 seconds
8-12 seconds
10-14 seconds
60-120 seconds
If ambient temperature greater
than 35°C (95°F): Immediate
60 seconds (after pulling probe
from probe well)
70 seconds (after measurement
site insertion)
± 0,1 °C (± 0,2 °F)
± 0,1 °C (± 0,2 °F)
± 0,2 °C (± 0,3 °F)
In standard predict mode,
thermometer accuracy meets
ISO 80601-2-56
A Standard prediction mode
reading and a direct mode reading
will differ by less than ±0.2°C
(±0.4°F) on 98% of tested patients.
Replaceable
For probe integrity/safety checks,
refer to qualified Technician or
return to factory.
Removable, replaceable,
washable, ABS 94 UL V0
Thermometer case: Flame
retardant ABS 94 UL V0 Isolation
chamber: Flame retardant
ABS 94 UL V0
Probe handle: Flame retardant
polycarbonate-polyester blend
Probe shaft: Flame retardant
polyester
Probe cable: Polyurethane jacket
with TPE over mould
Tip: Aluminium
Externally powered equipment
Type B
Continuous, Wall-mounted
Not suitable for use in the
presence of flammable aesthetic
mixture with air, oxygen or
nitrous oxide
Transport and store this device
between temperatures of -5°C to
50°C (23°F to122°F) Relative
humidity range should be
between 10%-85% RH, non-
condensing
10 °C bis 40 °C (50 °F bis 104 °F)
800 hPa - 1060 hPa
IPX0
12 VAC 1000 mA
12 VAC 1000 mA

Publicidad

Tabla de contenido
loading

Tabla de contenido