1 - Description and functions
The DynaCity foot includes the following main components :
An upper carbon fiber blade that allows energy return.
A carbon fiber plantar blade that ensures a progressive and smooth step rollover.
An intermediate carbon fiber blade allowing to get flat foot very well.
The 3 blades are split to provide good inversion-eversion at forefoot.
A polyurethane foot shell (1A60020) which transmits the mechanical stresses between the shoe
and the composite structure.
A technical fiber sock limiting noises and wear and tear between the foot shell and the structure.
A malleolar clip that ensures a good assembly with the cosmetic foam.
Thanks to the combination of these components, the foot has a comfortable and flexible behaviour.
2- Indications, maximum weight and connecting compo-
nents
The Dynacity foot is designed exclusively for lower limb pros-
thesis.
It is appropriate for patients with low or moderate activity level
whose weight (carried load included) complies with the table
opposite. The maximum load has been tested according to
standard
ISO-22675.
The DynaCity foot is designed for a 10 mm heel height.
The foot is resistant to fresh water only, provided that the
recommendations of § 8 are respected.
It includes a male pyramid that allows the use of standard adapters. Thanks to these adapters, the heel height can be
modified from 0 to 20 mm.
3– Assembling (without the patient)
Place under the foot a wedge of the same height as the shoe heel. Fix the pyramid
receiver onto the foot, taking care that its axis be perpendicular to the ground (fig. 1).
4- Alignment (with the patient)
Place the foot so that the load line goes through
the posterior third of the foot (fig. 2).
Coxo femoral
joint centre
Knee joint centre
Load line :
≈ 12mm ahead of the
pyramid axis (±5mm)
Pyramid axis
1/3
1A510
Instructions for Use
Knee in full
extension
Fig. 2
2/3
Category Selection Guide
Size
45 to 59 60 to 74 75 to 90 91 to 107 108 to125
22
1
23
1
24
1
25
2
26
2
27
2
28
2
29
4
30
4
* Categories are not modules : not only the patient
weight, but also the foot size must be taken into
account to determine the category.
H Heel
Vertical axis going through
the socket centre, the
knee and the pyramid
Front view : the
axis goes bet-
ween the 1st and
the 2nd toes
1A51099-0818- Page 5/20
Patient weight (Kg)
1
2
2
1
2
2
1
2
2
2
3
3
2
3
3
3
3
4
3
3
4
4
4
5
4
4
5
Adapter axis
Fig. 1
Maximum build
height : 130mm
NA
NA
NA
4
4
4
4
5
5