Siemens 7ME3951-0TG20 Instrucciones De Manejo Resumidas página 13

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Operation
Switch On
Press the [On/Off] key to turn on the gauge. The gauge will consecutively display:
● Logo - Siemens
● Part Number (MLFB) and software version
The gauge will then enter into the measuring mode and display:
● 0.0mm/0.00mm (0.00in/0.000in depending on the last used.)
● Velocity rate and selected probe
Backlight ON/OFF
● Press the [Change Setting/Calibration] key, the backlight is ON.
● Press the [Change Setting/Calibration] key again, the backlight is OFF.
Measurement
There are three ways to go into the measuring mode:
1. Switch on the gauge
2. Press power key [On/OFF] to escape from MENU
3. Measurement
Calibration
When the gauge is initially switched ON or if an error value instead of 4.00mm
appears (when velocity rate is 5920 m/s), or if you take a measurement of a test
block attached on the gauge (where the correct value should be 4.00mm) or on
another standard test block, then the gauge needs to be calibrated.
To calibrate the gauge, press the [Calibration] key. The gauge will display the same
information listed above, operate according to that information and then the gauge
automatically calibrate.
Set-up
Press [Menu/Confirm] key. The gauge will display:
1. Measurement
2. Velocity Rate
3. Probe Setting
4. Resolution
5. Default
Thickness Gauge
4
A6X30021999-06, 03/2012

Publicidad

Tabla de contenido
loading

Tabla de contenido