Setting the brightness
Use the control dial on the HEINE
Position I: maximun brightness
Position 0: OFF
Hygienic Reprocessing
Instructions on hygienic reprocessing must be adhered to, based on national standards, laws and guidelines.
Classification according to KRINKO: noncritical
Spaulding Classification USA: noncritical
Allow the device to cool down before reprocessing.
Before cleaning, disconnect the device from the power source.
In the event of suspected contamination, carry out hygienic preparation of the instrument.
The described cleaning and disinfection measures do not replace the specific rules applicable for the
establishment.
HEINE Optotechnik only approves the agents and procedures mentioned below.
Cleaning and disinfection may only be carried out by personnel with sufficient hygienic knowledge.
Observe the instructions of the manufacturer of the reprocessing media.
Do not use spray or immersion disinfection, dripping wet or heavily foaming tissues, automated or ultrasonic
reprocessing.
Procedure
Clean and disinfect the HEINE SIGMA
disinfect).
Recommended agents:
Cleaning agent: enzymatic (e.g. neodisher
Disinfectant agent: alcoholic (e.g. Incides
Changing the light source
With the HEINE SIGMA
Maintenance and Service
The instruments do not require maintenance or service.
We as manufacturer do not stipulate a safety inspection in accordance with § 6 Medical Device Operator
Ordinance (MPBetreibV), ref. attachment 1, for the HEINE product.
The expected operational life is 7 years.
General Notes and Warnings
Check the correct operation of the device before use! Do not use the device if there are visible signs
of damage.
Do not use the device in fire- or explosive risk area (e.g. oxygen saturated or anesthetic environments).
Do not modify the device.
Use only original HEINE spare parts, accessories and power sources.
Repairs shall only be carried out by qualified persons.
Do not use the device outside.
General Notes
The warranty for the entire product is invalidated if non-genuine HEINE products or non-original parts are
used and if repairs or modifications are made to the device by persons not authorized by HEINE. For more
information, please visit www.heine.com.
mPack mini to adjust the brightness of the HEINE SIGMA
®
250 (M2) and the HEINE
®
MediClean)
®
N) or agent hydrogen peroxide (e.g. PREempt
®
250 (M2) the LED cannot be changed.
®
MED 113379