Application Modes - Uline Ohaus Adventurer H-5276 Manual Del Usuario

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

SPAN CALIBRATION (SINGLE POINT)
Span calibration uses two calibration points, one at zero
load and the other at specified full load (span).
NOTE: Factory default settings are shown in bold.
With the balance turned on and no load on the pan,
touch SPAN CALIBRATION to initiate the procedure.
Additional calibration values to be used are shown on
the display. The best accuracy is achieved using the
mass closest to the full span value.
NOTE: To change the span calibration point,
touch the alternate weight shown on the display.
Follow the screen instructions and place the
specified calibration weight on the scale when
prompted to do so. When complete, the display
shows the span calibration status and returns to
the current application.
LINEARITY CALIBRATION (MULTI-POINT)
Linearity calibration uses three calibration points, one
at zero load and the others at specified loads. With
no load on the scale, press LINEARITY CALIBRATION to
begin the process. The balance captures the zero point,
and then prompts for the next weight. Continue to follow
the instructions on the display until the calibration is
completed.
NOTE: To cancel at any time, press CANCEL.
CALIBRATION TEST
Use Calibration Test to compare a known calibration
weight against the stored span calibration data.
With no load on the scale, press CAL TEST to begin the
process. The balance captures the zero point, then
prompts for the span value.
The display shows status, followed by the difference
between the current calibration weight and the stored
calibration data.
ZERO
Press ZERO to zero the balance.
TARING
With no load on the pan, press ZERO to set the display
to zero. Place an empty container on the pan and press
TARE. Add material to the container and its net weight
is displayed. Remove container and container's weight
appears as a negative number. Press TARE to clear.
PAGE 4 OF 54
OPERATION CONTINUED

APPLICATION MODES

Touch the top left application field (weighing in the
example below) to select the application mode.
(See Figures 4-5)
Figure 4
Figure 5
Use this application to determine the weight of items in
the selected unit of measure.
1.
2. Press TARE or ZERO if necessary to begin.
3. Place objects on the pan to display the weight.
4. The resulting value is displayed in the main display
Weighing
>0<
Gross:
Tare:
Settings
Figure 6
Weighing
639.77
Gross:
639.77 g
Tare:
0.00 g
Item
Sample
Settings
Name
Application Modes
Weighing
Parts Counting
Density
Determination
Check Weighing
Back
NOTE: Before using any application, be sure
the balance has been leveled and calibrated.
WEIGHING
In the upper left portion of the home screen,
(See Figure 6) select WEIGHING (this application is
the default).
When stable, the * appears.
line in the active unit of measure.
Max 4200 g
0.00
0.00 g
Capacity Guide
0.00 g
Item
Sample
Save to
Name
USB
Max 4200 g
d=0.01 g
g
Capacity Guide
4200 g
15%
Save to
Menu
USB
Percent Weighing
Dynamic Weighing
Display Hold
More...
Exit
The WEIGHING
d=0.01 g
home screen
Main Display Line
Touch g to
g
change unit
4200 g
Reference Fields
0%
Application
Menu
Buttons
0521 IH-5276

Publicidad

Tabla de contenido
loading

Tabla de contenido