A first press of the DISP key results in the last entered dispense volume being displayed. The VOLUME annunciator
will illuminate and flash. The INC/DEC ( ,
START/STOP key then initiates delivery of the set volume. The amount remaining to be dispensed will be displayed
during countdown. Resetting cumulative dispensed volume can be done by pressing START/STOP while holding
TOTAL VOLUME. Display flashes "999999" when cumulative volume is greater than 999999. The dispense function
is exited by pressing any key except Increment ( ), Decrement ( ), DISP, or START/STOP.
A second press of the DISP key causes the COPY annunciator to illuminate and flash. The START/STOP key is then
used to deliver the desired volume without the need to know the volume in specific units. A third press of the DISP
key enters the volume dispensed. The COPY annunciator stops flashing. The START/STOP key is then used to
initiate delivery of the copied volume. The number of copies dispensed is displayed after each dispense. The
maximum number of copies is 999999. The START/STOP key is used to pause the copy dispense during dispensing;
copy dispense can then be continued using the START/STOP key.
A fourth press of the DISP key results in the last entered dispense interval being displayed. The SECONDS
annunciator will illuminate and flash. The INC/DEC ( ,
from 1 to 999999 seconds. The START/STOP key then initiates delivery for the set time interval. The remaining time
will be displayed during countdown. Pressing the DISP key a fifth time exits this mode.

Keypad Lockout Enable/Disable

Press and hold FLOW. After five (5) seconds, display will change to all dashes. Then, while holding FLOW, press
PRIME five (5) times.
Selectable input (0–20 mA, 4–20 mA, 0–10V DC )
±0.5% linearity control
±50V common mode range
START/STOP; CW/CCW; PRIME via contact closure

Remote Control Setup

1. Place the power switch in the off position.
CAUTION:
Power must be turned off before connecting the external remote control cable to prevent
damage to the drive.
2. Connect the cable from the external remote control to the mating receptacle on the rear panel.
3. Select type of remote control input and output required as follows:
a. Press and hold the MODE control while turning the power switch to the on position. After two seconds,
release the MODE control. The initial display will show: "inP". After two seconds, the display shows either
0–20 or 4–20.
NOTE:
Press the UP ( , increment) or DOWN ( , decrement) arrow keys to select between 4–20 and 0–20 for
current loop control.
b. Press the MODE control again. The initial display will show: "out". After two seconds, the display shows either
0–10 or 0–20 and 4–20.
NOTE:
Press the UP ( , increment) or DOWN ( , decrement) arrow keys to select between 4–20 or 0–20 and
0–10 for current loop or voltage output.
4. Press the MODE control to select mode of operation. The LEDs indicate the selected mode. Select either mA or V.
NOTE:
If using only remote START/STOP, PRIME and/or CW/CCW, the MODE control can be set to any of the
three positions.

DISPENSE/COPY

) keys are used to change the dispense volume, if desired. The
) keys are used to change the dispense time, if desired,

REMOTE CONTROL