CALIBRATION

Calibration
This scale has been factory calibrated, and does not require calibration prior to use. If required, the
scale can be calibrated. Calibration of your 498KL scale is performed using kilograms (KG) or pounds (LB),
according to the unit of measure used upon entering into the calibration path. When calibrating in pounds, a
total of 300 lbs is required, applied in increments of 100 lb, 200 lb and 300 lb. If calibrating in kilograms, a total
of 120 kg is required, applied in increments of 40 kg, 80 kg and 120 kg.
Note: Only weights certified and traceable to national standards are to be used for calibration
procedures.
Pressing and holding the LB/KG and ZERO buttons at the same time, press
the ON/OFF button to turn on the scale. Release all three buttons when
"CAL" appears on the display. Numbers will appear indicating that you can
begin calibration.
When calibration numbers stabilize, press the ZERO button. "C-100" (for LB
mode) or "C-40" (for KG mode) will flash on the display, followed by "0".
Place 100 lb or 40 kg on the scale. Numbers will appear on the display.
When calibration numbers stabilize, press the ZERO button. "C-200" (for LB
mode) or "C-80" (for KG mode) will appear on the display, followed by "0" or
"1".
Place an additional 100 lbs or 40 kg on the scale, for a total of 200 lbs or 80
kg. Numbers will appear on the display. When calibration numbers stabilize,
press the ZERO button. "C-300" (for LB mode) or "C-120" (for KG mode) will
appear on the display, followed by "0" or "1".
Place an additional 100 lbs or 40 kg on the scale, for a total of 300 lbs or
120 kg. Numbers will appear on the display. When calibration numbers
stabilize, press the ZERO button. "END" will appear on the display, followed
by "OFF". The scale will automatically shut off.
.
ACTION
16
DISPLAY SHOWS
"CAL" followed by
numbers
Numbers followed
by "C-100" (LB
mode) or
"C-40" (KG mode)
then "0" or "1".
Numbers followed
by "C-200" (LB
mode) or
"C-80" (KG mode)
then "0" or "1".
Numbers followed
by "C-300" (LB
mode) or
"C-120" (KG
mode) then "0"
or "1".
"XXXXX" followed
by "END" then
"OFF"