Hach ORBISPHERE 410 Manual Básico Del Usuario página 3

Ocultar thumbs Ver también para ORBISPHERE 410:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 70
Table of Contents
1 Specifications
on page 3
2 Expanded manual version
3 General information
on page 4
4 Installation
on page 5
Section 1 Specifications
Specifications are subject to change without notice.
Specification
Ambient temperature
Storage temperature
Operating humidity
Operating altitude
EMC requirements
Korean registration
CE compliance
Safety rating
Enclosure ratings
Power supply
Analog current output
version on the
measurement board
Analog voltage output
version on the
measurement board
Measurement alarm
relays on the
measurement board
on page 4
Details
–5 to 50°C (23 to 122°F)
–20 to 70°C (–4 to 158°F)
0 to 95% non-condensing relative humidity
From 0 to 2,000 m. (6,550 ft.) above sea level
EN61326-1: EMC Directive
Note: The wall mount instrument is a Class A product. In a domestic environment this product may
cause radio interference in which case the user may be required to take adequate measures.
User Guidance for EMC Class A Equipment
업무용을 위한 EMC 등급 A 장치에 대한
사용자 지침
사용자안내문
A 급 기기 ( 업무용 방송통신기자재 )
이 기기는 업무용 (A 급 ) 전자파적합기기로서 판매자 또는 사용자는 이 점을 주의하시기
바라며 , 가정외의 지역에서 사용하는 것을 목적으로 합니다.
EN61010-1: LVD Directive
ETL, conforming to UL 61010-1 and CSA 22.2 No. 61010-1
IP 65; Totally protected against dust; Protected against low pressure jets of water from all
directions.
NEMA 4X (wall mount only); Totally protected against dust; Protected against pressure
jets of water from all directions.
Enclosure rating does not apply to external power supply for benchtop instruments.
Universal 100 VAC to 240 VAC @ 50/60Hz - 40VA; 10 to 30 VDC - 30W
4-20 mA (default) or 0-20 mA (configuration with software);
3 configurable outputs; Maximum load: 500 ohm; Sensitivity: 20μA;
Accuracy: ± 0.5% (between operating temperature limits)
0- 5 V output (hardware option);
3 configurable outputs; Minimum load: 10 KOhm; Sensitivity: 5 mV;
Accuracy: ± 0.5% (between operating temperature limits)
Three alarm relays; 1A-30 VAC or 0.5A-50 VDC on a resistance load
Configurable to Normally Open [NO] or Normally Closed [NC] contacts by changing the
jumper positions
Potential Electrocution Hazard. Connect only safety low voltage < 33 VAC RMS
5 User interface
on page 14
6 Start up
on page 14
7 Operation
on page 15
8 Maintenance
on page 24
W A R N I N G
W A R N I N G
English 3

Publicidad

Tabla de contenido
loading

Tabla de contenido