NSK Varios Combi Pro Manual De Instrucciones página 4

Ocultar thumbs Ver también para Varios Combi Pro:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 122
<Perio Mode>
• When using Perio Mode, pay attention not to cause emphysema as air containing powder is sprayed into periodontal pockets. If emphysema
is caused by treatment using Perio Mode, stop using this product immediately and seek medical attention.
• Do not use on the following patients. Using on periodontal pockets for these patients may result in bacillemia.
– Patients with endocardial inflammation
– Patients who have a contagious illness or disease
– Patients with an immune deficiency disease (neutropenia, granulocytopenia, agranulocytosis, diabetes, hemophilia)
– Patients undergoing irradiation treatment, chemotherapy, or antibiotic treatment
• Do not use on the following patients. (Excluding the using without the Nozzle Tip)
– Those with damage or abnormality in the oral cavity.
– Those with hyperemia, hemorrhage or inflammation in the oral cavity.
– Those that tend to develop inflammation or sores in the oral mucous membrane.
• Use this product only for maintenance after basic periodontal treatment. Never use it for other purposes.
• Use Perio Mode only on periodontal pockets that meet the following conditions:
– Periodontal pockets caused by gum disease
– Periodontal pockets of less than 6mm that have undergone basic periodontal treatment
• Check the depth of periodontal pockets with a measurement probe or photography before use.
• Use the designated cleaning powder only. Use of other powders may result in malfunction, substandard performance or damage to the
substantia ossea and dentine, or may cause emphysema. NSK cannot accept any liability for issues that occur as a result of the use of
powder other than those designated.
• Nozzle Tip for Perio Mode included in the product package is a sterile item. When using these products, follow the instructions below.
– Nozzle Tip is a single-use item. Do not reuse nor resterillize it as product breakage or infection could occur.
– Check before use that the package is NOT open nor damaged. If the package is open or damaged, do not use that Nozzle Tip as it may be
damaged or contaminated.
– Observe the use-by date written on the Nozzle Tip package label. Do not use expired products as product breakage or infection could
occur.
<Ultrasonic Mode and Powder Mode>
• Read this Operation Manual before use to fully understand the product functions and file for future reference.
• When operating the product always consider the safety of the patient.
• The end user shall be responsible for any judgment that relates to the application of this product to a patient.
• Users are responsible for the operational control, maintenance and continual inspection of this product.
• This device is for indoor use only.
• Keep the Control Unit on a level surface.
• Do not attempt to disassemble the product nor tamper with the mechanism except as recommended by NSK in this Operation Manual.
• Do not allow any impact on to the product. Do not drop the product. The Glass Rod and Handpiece may be damaged if the Ultrasonic
Handpiece is dropped.
• Operators and all others in the area must wear eye protection and a mask while using this product.
• Should the product function abnormally, cease operation immediately and contact your Authorized NSK Dealer.
• Push the Handpiece securely and fully onto the Handpiece Cord or Handpiece Hose when connecting.
• Do not autoclave the Handpiece Cord or the Handpiece Hose. Use alcohol to disinfect the Cord or Hose after use.
• Do not use high acid water or sterilizing solutions to wipe, immerse or clean the product.
• This product is not delivered in a sterile condition, except for the Nozzle Tip (option) and must be autoclaved prior to use. Refer to "9-3
Sterilization" for details.
• Perform regular function and maintenance checks.
• If the product is not used for a long period check it is functioning correctly before using on a patient.
• To avoid clinical downtime it is recommended that a spare be kept on hand in case of a breakdown during surgery.
• Always replace the fuse with the same specified rating.
• This product is rated Medical Electrical equipment. EMC (Electromagnetic compatibility) is described in the documentation included.
• Installation and use of this product requires special precautions regarding EMC according to the EMC information.
• Portable and mobile RF communications equipment can affect Medical Electrical equipment. Do not use RF equipment near the product.
• The use of ACCESSORIES such as handpieces and cables, other than those specified by the manufacturer, with the exception of handpieces
and cables sold by the manufacturer of this product as replacement parts for internal components, may result in increased EMISSIONS or
decreased IMMUNITY of this product.
• This product should not be used adjacent to, or stacked with, other equipment. If adjacent or stacked use is necessary, this product should
be observed to verify normal operation in the configuration in which it will be used.
WARNING
CAUTION
3

Publicidad

Tabla de contenido

Solución de problemas

loading

Tabla de contenido