6.2 Control panel

The control panel allows function selection and shows the autoclave alarm status.
Once the unit is plugged in, put it on, by switching on the main switch. The unit goes
7
to stand by status (ON).
15 Pressure display
It displays chamber pressure in real-time during the cycle.
16 Temperature
It displays chamber temperature during a cycle. The units is per Celsius degrees
(ºC).
17 State displays/Error codes/Time
It displays all the state of autoclave and the error codes when alarm goes off.
You may correct the errors accordingly the error list.
18 SELECT key
Programme selection key. It is used for selecting a temperature (121 or 134) or
the textile mode (B
It can also be used to make a washing cycle.
19 PACKAGE key
It is used to select wrapped or unwrapped instrumental.
20 TEST key
You can select the Bowie Dick test or the vacuum test.
21 STEP key
This is a post cycle selection key. By pressing this key, the autoclave keeps the
21
chamber warm, and thus it saves time in the following cycle.
If you want to cancel this operation, press this key again. The indicator will turn
off.
22
22 Quick DRY key
This key short drying time (1 or 5 min) is selected. It is only recommended in
23
processes with not wrapped material.
23 START/STOP key
This key is used to start programmes and abort them manually. If you press it
for more that 5" during the operation, it aborts the cycle and starts a 2-minute
drying cycle.
Autovía A2 Km 585.1 Abrera 08630 (Barcelona) España
e-mail: selecta@jpselecta.es - website: http://www.jpselecta.es
JULY/2018
(It can be modified without notice)
).
+
Page: 22
Tel 34 937 700 877 Fax 34 937 702 362

Hide quick links:

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

4001737

Tabla de contenido