You have made the right decision by choosing a measuring instru-
ment from the Testo stick product line. If handled properly you will
have many years of reliable and accurate measured results.
Lower part of probe pipe can be
turned by 90° enabling display to be
read while at the same time
observing the flow direction (see
arrow on protection cup).
Universal joint, can
be bent by 180°.
Technical data
Measuring range m/s: ....................0 to 5m/s at -20 to 0 °C;
0 to 10 m/s at 0 to +50 °C
3
Measuring range m
/h: ............................0 to 99.990 m
Measuring range °C: ....................................-20 to +50 °C
Resolution: ................................................0.01 m/s; 0.1 °C
Accuracy: (system accuracy at calibration temp. +25°C)
±5 % of m.v. ±0.10 m/s (to 2 m/s);
±5% of m.v. ±0.3 m/s (over 2 m/s)
Ambient temperature: ......................................0 to +50 °C
Storage temperature: ....................................-20 to +70 °C
Battery type: ..............................................3 x micro (AAA)
Battery lifetime: ..............................................Approx. 20 h
(approx. 750 measurements lasting 2 minutes each)
Probe stem: ............................................Ø 12 mm/16 mm /
Length: 300 mm
Warranty:..................................................................2 years
Please read before using instrument
• Observe flow direction • Adhere to sensor measuring
ranges • Do not exceed maximum storage and operating
temperatures (e.g. protect measuring instrument from
direct sunlight)!
Inexpert handling cancels your warranty.
Measuring instruments conform to EN 50 081-1 +
EN 50 082-1
Instruction manual
testo 405-V1
The velocity sensor is protected by a
quick twist of the bottom of the probe
stem
the protective cap is closed.
Should only be left open for the
duration of the measurement.
Ø 16 mm
180°
3
/h
± 0.5 °C
Befestigungs-Clip
Holder for positioning stick in
duct e.g. in air con duct, see
Fig. opposite.
Switching on
On
Press button once:
segment test follows.
Velocity stick
The velocity value is shown on the display after
the segment test.
Switching off
On
Ø 12 mm
Keep button pressed for 3 seconds.
Auto Off function
If no button has been pressed in 5 minutes,
the instrument switches itself off automatically.
Changing the battery
The battery has to be changed if the
symbol appears in the display during the
measurement.
Turn the bayonet coupling on the head part to
the left and then press head back. Remove
empty battery and replace with new battery -
Observe polarisation - and close coupling again.
Swivel head.
The display can always
be read.
Flow direction
An arrow on the protective cap shows the flow
direction in which the stick was calibrated and
the best readings obtained.
Changing the parameters
Press button repeatedly:
On
m/s
On
2. 2 2
On
45. 8
m
3
/h x 10
°
C
2220
Set duct cross-section
On
Keep On button pressed when switching on.
2
m
1st position flashes
0. 0 00
On
m
2
1. 0 00
On
m
2
On
m
2. 0 00
3. 0 00
- until the required digit is shown.
Wait 3 seconds and the digit will apply. The
cursor then changes to the next position.
m
2
2nd position flashes
3. 0 00
On
m
2
On
2
m
3. 1 00
3. 2 00
etc.
- until the required digit is reached.
Wait 3 seconds and the digit will apply. The
cursor then changes to the next position.
Any digit between 0 and 9 can be set at each
position .....
= duct cross-section in m
2
etc.
2