C M=0.1000 A=0.0000
Save Calibration? Yes m
13 TDL GPro™ 500
For Display setup, analog outputs and setpoints see the section "General Setup"
h
TDL Installation (PATH: Quick Setup / TDL / Installation)
20.9
B
%V O2
20.9
B
%Trm
MENU
Quick Setup
0.0
B
%V G1
0.0
B
%V G2
Transmission: 000.00%
NSL:000%
TDL Commissioning (PATH: Quick Setup / TDL / Commissioning)
20.9
B
%V O2
20.9
B
%Trm
O2–TDL
Commissioning
22
After the calibration of the cell multiplier or slope calibration factor "M" i.e. cell con-
stant and the Adder or offset calibration factor "A" are displayed.
Select YES to save the calibration values and the successful C alibration is confirmed
on the display.
The user gets the message RE-INSTALL SENSOR and "Press [ENTER]" on the display.
After pressing [ENTER] the M400 returns to the measuring mode.
Note: TDL is calibrated at the factory and does not require calibration at installation
& Startup.
While in measurement mode press the key [MENU]. Press the m or . key to select
the TDL and then the Installation menu item.
u
In this mode, the current live value of the % transmission and the Noise Signal
Level (NSL) are displayed during 5 minutes until it automatically returns to the Mea-
surement mode. Best performance of the TDL is obtained when the % transmission
value is maximized (50% or more) and the NSL is minimized (50 or less). Use these
u
values to rotate the blue sensor head attached with a loose clamp connection on to
the probe until the maximum transmission is found. In this position, and secure the
blue sensor head into position and tighten the clamp.
While in measurement mode press the key [MENU] . Press the m or . key to select
the TDL and then the Commissioning menu item.
First, select the type of pressure compensation selected:
– External: c urrent external pressure value coming from a pressure transducer
u
of 4...20 mA analog output