Guidance and manufacturer's declaration – Immunity Emissions
The SUPERVEGA EVO SUCTION UNIT is intended for use in the electromagnetic environment specified below.
The customers or the user of the SUPERVEGA EVO SUCTION UNIT should make sure that it's used in such an envi-
ronment.
Immunity Test
Level indicated by the
EN 60601-1-2
Conducted
3Vrms 150kHz
Immunity
to 80Mhz
EN 61000-4-6
(for non life-supporting
devices)
Radiated
10V/m 80MHz
Immunity
to 2.7GHz
EN 61000-4-3
(for non life-supporting
devices)
Note 1: At 80 MHz and 800 MHz the interval with the highest frequency is applied
Note 2: These guide lines may not be applicable in all situations. The electro-magnetic propagation is influenced by the
absorption and by reflection from buildings, objects and people.
a) The field intensity for fixed transmitters such as the base stations for radiotelephones (mobile and cordless) and terres-
trial mobile radio, amateur radio devices, radio AM and FM transmitters and TV transmitters can not be theoretically and
accurately foreseen. To establish an electro-magnetic environment generated by fixed RF transmitters, an electro-magnet-
ic study of the site should be considered. If the field intensity measured in the place where the device will be used surpass-
es the above mentioned applicable level of conformity, the normal functioning of the device should be monitored. If abnor-
mal performance arises, additional measures such as changing the device's direction or positioning may be necessary.
b) The field intensity on an interval frequency of 150 kHz to 80 MHz should be less than 3 V/m.
Compliance
Electromagnetic environment - guidance
Level
V1 = 3 V rms
The portable and mobile RF communication devices,
including cables, must not be used closer to the SU-
PERVEGA EVO SUCTION UNIT device, than the sep-
aration distance calculated by the equation applicable
to the transmitter frequency.
Recommended separation distance
3,5
d=
E1 = 10 V / m
V
12
d=
E
23
d=
E
Where P is the maximum nominal output voltage of the
transmitter in Watt (W) depending on the manufactur-
er of the transmitter and the recommended separation
distance in metres (m). The intensity of the field from
the fixed RF transmitters, as determined by an elec-
tro-magnetic study of the site
level of conformity of each frequency interval
It is possible to check for interference in
proximity to devices identified by the fol-
lowing symbol:
ENGLISH
√P
1
from 80 MHz to 800MHz
√P
1
from 800 MHz to 2.7 GHz
√P
1
, could be lower than the
a)
.
b)