PROGRAMMING
SETTINGS
TASK
PROGRAM 0
PRO.: SOL. 2
TEMP.: 134
T.EST.: 4m
T.SEC.: 10m
ACTUAL: +0.00 bar 24ºC
Drying time
Shows running
program status
PROGRAM 0
-0.70 bar
24 ºC
TIME: 00:00:06
S2 - 134ºC - EST 4' SEC 10'
STERILIZATION
CORRECT
DOOR OPEN IS ENABLED
ACTUAL: +0.00BAR 23ºC
J
.P. SELECTA s.a.u.
8.5 Load
Open the lid. Check tank water level.
Load the previously cleaned and settled material baskets.
Check that the seal (boiler top) is in place.
If it is installed and if it is necessary, place the heart probe in the chosen refe-
rence point.
Close the lid.

8.6 Select Program

Select the suitable program for sterilization and drying the load.
See chapter 8.4 to set program cycle.
C
O
RUN
HH:MM:SS DD/MM/YY
8.6.1 Application of the cycle
STEP: WATER
During the cycle (also called program) autoclave screen will indicate the phase
that is being made.
Phases are:
WATER:
STOP
VACUUM:
HH:MM:SS DD/MM/YY
HEAT:
STERILE:
EMPTY:
UNPRESS:
OK:
FAIL:
OK
This is the "End of cycle" message.
HH:MM:SS DD/MM/YY
Autovía A-2 Km 585.1 Abrera 08630 (Barcelona) España
e-mail: selecta@jpselecta.es - website: http://www.jpselecta.es
Ene 2020
(Sujetas a modificaciones sin previo aviso)
Water inlet to the boiler.
Air extraction.
Steam generation.
Sterilization of the load.
Draining the boiler steam and water.
Depressurization boiler.
Sterilization conditions have been achieved according to
cycle parameters.
Not achieved all the conditions for the sterilization.
Tel 34 93 770 08 77 Fax 34 93 770 23 62