INSTRUCTION MANUAL CODE 80311 REV E
J
.P. SELECTA s.a.u.
Jan 2020
12
Maintenance
Before removing the lid, disconnect the machine from the
mains.
The control board should only be changed by a technical ser-
vice or authorized personnel.
CLEANING:
•
Wall panels and stainless steel: Use steel cleaning home products.
•
Limescale deposits: Use home cleaning products.
•
TFT Screen: use a cloth and a glass-cleaning liquid
•
Plastic lids: use a cloth and a glass-cleaning liquid
Monthly operations:
•
Clean front bottom drainage filter.
•
Change re-used water from water reservoir.
•
Record results from vacuum test and Bowie Dick test.
Three months operations:
•
Change air entry filter (autoclave rear side)
Yearly operations:
•
Check and contrast temperature autoclave against traceable
temperature standard.
•
Check safety valve operation.
•
ATTENTION: this operation should be run for qualified personnel.
Dangerous contact with hot steam.
•
Visually check the state of electrical wiring, ground and mains.
Autovía A-2 Km 585.1 Abrera 08630 (Barcelona) España
e-mail: selecta@jpselecta.es - website: http://www.jpselecta.es
(It can be modified without notice)
Tel 34 93 770 08 77 Fax 34 93 770 23 62
Page: 42