15ml glass tube
Bottom rubber
 
Error code 
Code Meaning 
E1 
Unbalance protection 
 
 
E2 
Over speed 
 
 
E3 
Door cover protection 
E6 
Over current
E7 
No speed measurement 
E11 
The set speed can't be 
reached 
 
 
No speed measurement of motor after starting, please turn the motor manually. Observe whether the LED beside the speed measuring interface 
flashes, If it does not flash, the speed sensor may be damaged. If flashing, contact after‐sales service. 
 
J
.P. SELECTA s.a.u.

9. PLACING TUBES

Bucket

10. ERRORS TABLE

Malfunction Solutions ‐ Cencom II
Reason
Unbalanced protection input switch closed 
 
Speed measurement interference or speed rise too fast Check Speed measuring wire or reduce the speed up 
Door cover input signal wire is not connected properly 
or Door cover switch fault  
Current over rated value
No speed signal input
 
The speed sensor damaged or wire connection loose 
Insufficient motor power
External voltage too low  
Ctra. NII Km 585.1 Abrera 08630 (Barcelona) España
e-mail: selecta@jpselecta.es - http://www.jpselecta.es
January/2020
(It can be modified without notice)
15ml FALCON tube
Resolution
Check imbalance detection switch 
Unplug the unbalanced protection wire, no alarming occurred 
Check the door cover input signal wire and switch 
Speed up too fast under heavy load, reduce speed increasing
Drive board failure  
External voltage too high 
Check speed sensor and make sure the wire connection is good
 
There are many reasons for the input of no speed measurement signal 
 
Analyze the cause separately 
Ensure load and control system power match
Ensure the power supply is normal  
 