Accuracy Table
AC current (Measurement value/MAX/MIN/AVG)
(Accuracy
Range
guarantee range)
600.0 A
(1.0 A to 600.0 A)*
2000 A
(10 A to 1800 A)
(1801 A to 2000 A)
Auto range movement threshold: 6000 counts or more for upper range, 540 counts or less for lower range.
* For 30.0 A or less, add 0.5 A to the measurement accuracy.
52
Accuracy guarantee
Resolution
frequency range
0.1 A
10 Hz≤f<45 Hz
45 Hz≤f≤66 Hz
66 Hz<f≤1 kHz
1 A
10 Hz≤f<45 Hz
45 Hz≤f≤66 Hz
66 Hz<f≤1 kHz
1 A
10 Hz≤f<45 Hz
45 Hz≤f≤66 Hz
66 Hz<f≤1 kHz
HIOKI CM4371A976-00
Measurement accuracy
Filter off
Filter on
±1.8% rdg.±0.5 A
±2.3% rdg.±0.5 A
±1.3% rdg.±0.3 A
±1.8% rdg.±0.3 A
±2.0% rdg.±0.5 A
-
±1.8% rdg.±5 A
±2.3% rdg.±5 A
±1.3% rdg.±3 A
±1.8% rdg.±3 A
±2.0% rdg.±5 A
-
±2.8% rdg.±5 A
±3.3% rdg.±5 A
±2.3% rdg.±3 A
±2.8% rdg.±3 A
-
-