Rush Current (INRUSH) of AC+DC current
Range
(Accuracy guarantee range) Resolution
2000 A
(100 A to 1800 A)
(1801 A to 2000 A)
Rush Current (INRUSH peak value) of AC+DC current
Range
(Accuracy guarantee range)
2000 A
(±100 A to ±2300 A)
(±2310 A to ±2840 A)
9
AC voltage
CMRR:
Zero-display range:
Coupling type:
Crest factor:
Peak detection time width:
Overload protection:
Transient overvoltage:
1 A
1 A
Resolution
10 A
10 A
-60 dB or more (DC, 50 Hz/60 Hz, 1 k
5 counts or less
AC coupling
For 4000 counts or less, 3
For greater than 4000 counts and less than or equal to 6000 counts, 2
For 850 V or less, 2 (1000 V range only)
For greater than 850 V and less than or equal to 1000 V, 1.7
1 ms or more (Filter off)
Lower of 1870 V DC/1100 V AC or 2×10
8000 V
HIOKI CM4371A976-00
Accuracy guarantee
frequency range
DC, 20 Hz≤f≤500 Hz
DC, 20 Hz≤f≤66 Hz
Accuracy guarantee
frequency range
DC, 20 Hz≤f≤500 Hz
DC, 20 Hz≤f≤66 Hz
Ω
unbalance)
7
V • Hz (energized for 1 minute)
Accuracy Table
Measurement accuracy
±3.3% rdg.±13 A
±5.0% rdg.±13 A
Measurement accuracy
±6.0% rdg.±100 A
±8.0% rdg.±100 A
57