[12] HF-connection
Connection for the HF-cable (see paragraph 6.8 of this manual).
[13] Connection mains cable
Connect the Radarmed to a grounded safety socket of a voltage corresponding to the data mentioned on
the type plate of the unit.
[14] Connection for potential equalization cable
A special cable can be supplied for areas where local regulations require connection of a potential
equalization cable.
[15] Type plate
The type plate gives details of the model, series- and serial number (for service, warranty etc.) as well as
ratings such as mains voltage and power consumption.
8.2
Operation
8.2.1
Set up of the Radarmed 650+
Switching on
∙
Connect the radiator.
∙
Connect the HF-cable to the radiator and Radarmed.
∙
Switch on the unit using the On/off switch.
Immediately after switching on, the unit carries out a self test.
Choice of therapy mode
∙
Select continuous or pulsed mode
The illuminated LED indicates the selected mode.
Alignment of the radiator
∙
Align the radiator to the area which must be treated. Align
the radiator carefully so that other parts of the body are
minimal in the direction of radiation.
∙
Inform the patient.
For the alignment of the radiators take notice of the following
distances of the radiator to the body surface.
∙
Local field radiator approx. 10 cm.
∙
Longitudinal radiator approx. 5 cm.
∙
Large field radiator approx. 1 cm.
Treatment time
∙
Select the treatment time.
The set time is indicated on the display.
Power adjustment
∙
Set the power.
The set power is indicated on the display. At the end of the
treatment time the HF-power switches off.
∙
The treatment time runs down as soon as a power is set.
A flashing dot in the display indicates that the clock is
running.
.
A short beep every 5 seconds indicates microwave
energy is being produced.
Treatment starts immediately after operation of the power
setting selectors.