ELECTROMAGNETIC COMPATIBILITY (EMC)

• Guidance and manufacturer's declaration – electromagnetic emissions
The device is intended for use in the electromagnetic environments listed below, and should only be used in
such environments:
Emissions test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions IEC
61000-3-2
Voltage fluctuations/flicker
emissions IEC 61000-3-3
• Guidance and manufacturer's declaration – electromagnetic immunity
The device is intended for use in the electromagnetic environments listed below, and should only be used in
such environments:
Immunity test
IEC 60601 test level
Electrostatic
± 8 kV contact discharge
discharge (ESD)
IEC 61000-4-2
± 15 kV air discharge
Power frequency
30 A/m
(50 or 60 Hz)
50 or 60 Hz
magnetic field
IEC 61000-4-8
Electrical fast
± 2 kV for power supply
transient/burst
IEC 61000-4-4
± 1 kV for input/output lines
Surge
± 2 kV Power lines
IEC 61000-4-5
Interruptions and
0% UT; 0.5 cycle
voltage variations
At 0°, 45°, 90°, 135°, 180°,
on power supply
225°, 270°, and 315°
input lines
0% UT; 1 cycles
70% UT; 25/30 cycles
IEC 61000-4-11
0% UT; 250/300 cycles
Compliance
Group 1
Class B
Class A
Complies
lines
Electromagnetic environment – guidance
RF energy is used only to maintain device's operation. Therefore,
its RF emissions are so low that it's not likely to cause any
interference in nearby electronic equipment.
The device is suitable for use in all establishments, including
domestic establishments, and those directly connected to the
public low-voltage power supply network that supplies buildings
used for domestic purposes.
Compliance level
± 8 kV contact discharge
± 15 kV air discharge
3 A/m
50 or 60Hz
± 2 kV for power supply
lines
± 1 kV for input/output
lines
± 2 kV Power lines
0% UT; 0.5 cycle
At 0°, 45°, 90°, 135°, 180°,
225°, 270°, and 315°
0% UT; 1 cycles
70% UT; 25 cycles
0% UT; 250 cycles
27
Electromagnetic environment–
guidance
In air discharge testing, the climatic
conditions shall bw within the
following rages:
Temperature:15˚~35˚,
Humidity:30%~60%.
Power frequency magnetic fields
should be at levels characteristic of a
typical location in a typical commercial
or hospital environment.
Mains power quality should be that
of a typical commercial or hospital
environment.
Mains power quality should be that
of a typical commercial or hospital
environment..
Mains power quality should be that
of a typical commercial or hospital
environment. If the user of the device
requires continued operation during
power mains interruptions, it is
recommended that the device be
powered from an uninterruptible power
supply or a battery.