Electromagnetic Compatibility (Emc) - HoMedics HEALTH PLUS BPW-040-HP Manual De Instrucciones

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

ELECTROMAGNETIC COMPATIBILITY (EMC)

• Guidance and manufacturer's declaration – electromagnetic emissions
The device is intended for use in the electromagnetic environments listed
below, and should only be used in such environments:
Emissions test
Compliance
RF emissions CISPR 11
Group 1
RF emissions CISPR 11
Class B
Harmonic emissions IEC
Not
61000-3-2
Applicable
Voltage fluctuations/
Not
flicker emissions IEC
Applicable
61000-3-3
• Guidance and manufacturer's declaration – electromagnetic immunity
The device is intended for use in the electromagnetic environments listed
below, and should only be used in such environments:
Immunity test
IEC 60601 test level
Electrostatic
± 6 kV contact
discharge (ESD)
± 8 kV air
IEC 61000-4-2
Power frequency
3 A/m
(50/60 Hz)
magnetic field
IEC 61000-4-8
Electromagnetic environment
– guidance
RF energy is used only to maintain
device's operation. Therefore, its
RF emissions are so low that it's
not likely to cause any interference
in nearby electronic equipment.
The device is suitable for use
in all establishments, including
domestic establishments, and
those directly connected to the
public low-voltage power supply
network that supplies buildings
used for domestic purposes.
Complianc
Electromagnetic
level
environment–
guidance
± 6 kV contact
Floors should be
± 8 kV air
wood, concrete
or ceramic
tile. If floors
are covered
with synthetic
material, the
relative humidity
should be at
least 30 %.
3 A/m
Power frequency
magnetic fields
should be at
levels character-
istic of a typical
location in a
typical commer-
cial or hospital
environment.
27
• Recommended separation distances between portable and mobile RF
communication equipment and the device.
The device is intended for use in an electromagnetic environment where
radiated RF disturbances are under control. User can help prevent
electromagnetic interference by keeping the device at a minimum distance from
portable and mobile RF communications equipment (transmitters). Below table
details the maximum output power of transmitter:
Rated maximum
Separation distance according to frequency of
output power of
transmitter W
150 kHz to 80 MHz
Not Applicable
0.01
NA
0.1
NA
1
NA
10
NA
100
NA
For transmitters rated at a maximum output power not listed above,
the recommended separation distance d in metres (m) can be esti-
mated using the equation applicable to the frequency of the transmit-
ter, where P is the maximum output power rating of the transmitter in
watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the
higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electro-
magnetic propagation is affected by absorption and reflection from
structures, objects and people.
transmitter m
80 MHz to 800 MHz
800
MHz
d = 1.2
to 2.5
GHz
d = 2.3
0.12
0.23
0.38
0.73
1.2
3.8
12
28
2.3
7.3
23