ELECTROMAGNETIC COMPATIBILITY INFORMATION
Table 1 For all ME EQUIPMENT and ME SYSTEMS
Guidance and manufacture's declaration – electromagnetic emission
The [BPW-1005] is intended for use in the electromagnetic environment specified below.
The customer or the user of the [BPW-1005] should assure that it is used in such an environment.
Emission test
RF emissions
CISPR 11
RF emission
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Table 2 For all ME EQUIPMENT and ME SYSTEMS
Guidance and manufacturer's declaration - electromagnetic immunity
The [BPW-1005] is intended for use in the electromagnetic environment specified below.
The customer or the user of the [BPW-1005] should assure that it is used in such an environment.
Immunity test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
All manuals and user guides at all-guides.com
Compliance
The [BPW-1005] uses RF energy only for its internal
function. Therefore, its RF emissions are very low and are
Group 1
not likely to cause any interference in nearby electronic
equipment.
Class B
The [BPW-1005] is suitable for use in all establishments
Not applicable
other than domestic and those directly connected to the
public low-voltage power supply network that supplies
buildings used for domestic purposes.
Not applicable
IEC 60601 test level
±6 kV contact
±6 kV contact
±8 kV air
±8 kV air
3A/m
3A/m
12
Electromagnetic environment - guidance
Compliance level
environment - guidance
Floors should be wood,
concrete or ceramic tile.
If floors are covered with
synthetic material, the
relative humidity should be
at least 30%.
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment
Electromagnetic