2.4.2
Electrical Shock or Fire Hazard
No internal, operator-serviceable parts
Do not open unit, remove covers or attempt to repair the AED7000. All servicing must be
performed by qualified personnel.
Improper use can cause injury
The AED7000 contains an automatic disarm of the stored energy. If the operator has not
delivered the energy to a patient or a test load, an internal timer will disarm the stored
energy. This stored electrical energy can potentially cause death or injury if discharged
improperly. Follow all instructions in this User's Manual.
Do not immerse or expose the AED7000 to water or other liquids
Do not use the defibrillator if the unit has been immersed in liquid or if excessive conden-
sation is visible on the device.
Conductive parts should not contact other conductive parts including the earth
2.4.3
Improper Device Performance
Properly use electrode pads
Do not attempt to warm the electrodes with a heat source greater than 35°C (95°F). Do
not immerse or clean electrodes with alcohol or solvents. Do not perform chest compres-
sions (CPR) through electrodes. These actions may damage the electrode pads and cause
the unit to function improperly.
Use only M&B approved accessories
Do not use defibrillation pads, batteries, and other accessories not approved by M&B. Use
of unauthorized accessories may cause the device to operate improperly and provide false
measurements. Follow all labeling instructions on the defibrillation pads and the battery.
Do not repeatedly charge and discharge defibrillator in rapid succession
If a need for repetitive testing arises, wait at least 1 minute for every third discharge to
avoid damaging equipment
Improper maintenance can cause improper performance
Follow instructions in the Users Manual.
12 12