If an unpredictable situation occurs, the operator should press the ON/OFF button for 2 seconds to
shut down the AED7000 and internally discharge all the power to the machine
6. MAINTAINING THE SEMIAUTOMATIC EXTERNAL DEFIBRILLATOR
The maintaining, cleaning and Troubleshooting of the AED7000 will be described in this section:
• Maintaining
• Cleaning and disinfecting
• Troubleshooting
6.1
Maintenance
Maintenance is required for the optimal operation of the AED7000. Maintenance should be perfor-
med periodically; with weekly or monthly inspections recommended.
The frequency of inspection:
It is recommended that the AED7000 is inspected once a week or twice a month, but if use of the
AED7000 is more frequent, additional inspections should be performed.
The table below describes a general rule to perform the inspection
Frequency of use (n / month)
1 / month
2 ~ 3 / month
Infrequent using(1 ~ 2 a year)
Caution!
If the AED7000 should be used more than 4 times a month, we recommend
that the AED7000 should be inspected by authorized service personnel at
least twice a year
6.1.1
Inspection schedule
The general inspection schedules are described below. The operator should follow these schedules
when performing the inspection, as improper inspection may cause damage to the AED7000.
Check the status of indicator and the voice prompt.
If the AED7000 is started without electrode pads connection, the indicator of "Check pad" will be fli-
ckering and a voice prompt of "Check the electrode" will be broadcast. In the absence of either status
indicator above, contact the authorized service personnel for the necessary servicing or repair.
24
Inspection
Once a week
Twice a week
Once a month