Specification; Physical - Mach Power HC-AED7000 Manual Del Usuario

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Time out, analysis restart
Solution:
• If the operator doesn't press the shock button in 30 seconds after the analysis is finished, the
AED7000 will automatically disarm.
• Press the shock button in 30 seconds after the voice prompts of "press flashing shock button".
Press flashing shock button
6.3.5
Troubleshooting of battery
Problem description:
During the operation of the AED7000, the ON/OFF indicator is flickering.
Solution:
Shut down the AED7000 and replace the current battery with a new one.

7. SPECIFICATION

The Specification of physical, electronic isolation, defibrillator, battery, electromagnetic compatibility,
and rhythm recognition performance is described in this section:

• Physical

• Electronic isolation
• Defibrillator
• Battery
• Electromagnetic compatibility
• Rhythm recognition performance
7.1
Physical
Category
Dimensions
Weight
Operating Temperature
Operating Humidity
Altitude
Pressure
Storage Temperature
Storage Humidity
28
Specification
303 x 216 x 89 mm
2.0 kg
0°C -40°C
Relative humidity between 30% and 95% (non-con-
densing)
-91m a 4573m (-300 a 15,000 ft)
102.4kPa~57.2kPa (768mmHg~429mmHg)
-20°C a 50°C (without battery)
Up to 95% (without battery and non-condensing)

Hide quick links:

Publicidad

Tabla de contenido
loading

Tabla de contenido