2.5

Battery

2.5.1

Battery care

Use only M&B batteries
Use only M&B Part No. CR123A-4x 2 non-rechargeable lithium batteries. Use of any other
battery can damage the AED7000.
Always verify remaining capacity of a non-rechargeable battery after use
Check capacity and change if battery is low.
Replace battery at 5 years
Battery replacement at 5 years is recommended due to degradation of the battery chemi-
stry. Periodic maintenance and testing is highly recommended to ensure proper battery
performance.
2.6

Care and Storage

2.6.1

Cleaning and Sterilization

Clean and maintain the AED7000 according to instructions. See Chapter 4, Maintaining
the AED7000 Automatic External Defibrillator.
Do not clean the AED7000 with alcohol, ketone, or any flammable agent. Do not autocla-
ve the AED7000 or attempt to sterilize the AED7000 or any of its accessories.
2.6.2
Electrodes
Follow manufacturer's instructions for use of defibrillation electrodes
Improper use of defibrillation electrodes may cause the AED7000 to function improperly
or may cause skin burns. Do not use expired, dry electrodes. Do not reuse disposable
electrodes. Only use M&B approved electrodes.
Properly store and use defibrillation pads
Store electrodes in a cool, dry location (between -4° and 131°F or -20°C and 55°C). Do not
sterilize the pads, immerse, or clean the electrodes with alcohol or solvents.
2.6.3
Disposal
Disposal of device and its accessories
Please dispose of the device and its accessories which have exceeded their operating time
according to relevant regulations. If you have any question on device disposal or need
more detailed information, please contact us or our service centers.