The AED7000 takes 8~10seconds to complete analysis. Because the analyzing result depends on the
ECG waveform of the patient, which may be interrupted by any touching or moving the body of the
patient, any touch or movement may lead to an incorrect analysis result.
Warning!
Don't touch or move the patient during the analysis period.
There are two possible analysis results. If defibrillation is advised, the AED7000 will start the charging
procedure; otherwise, a recommendation of Cardiopulmonary resuscitation procedure (CPR) will
begin.
5.6
Charging procedure
If the AED7000 determines that the patient requires defibrillation, the charging procedure will start.
During the charging procedure, Caution indicator and analyzing indicator will be
flickering
Charging, do not touch the patient.
Warning!
Do not touch the patient during the charging period
First
energy
level
The AED7000 has three energy of defibrillation,as shown in the table below:
First energy level
Second energy level
Third energy level
The charging time depends on the energy of defibrillation
Second
2min CPR
energy
level
2min CPR
150J
150J
200J
Third
2min CPR
energy
level
21