Operating Temperature

10° - 40° C (50° - 104° F) 75% R.H. non-condensing
Packaged Storage and Transport Temperature
0° - 50° C (32° - 122° F) 90% R.H. non-condensing
Unpackaged Storage and Transport Temperature
Store between 0º – 50º C (32º F- 122º F) at < 90% RH (non-condensing); Excursions permitted to -25º C for up to 24 hours.
The pump requires 60 minutes to warm up to its normal operating temperature from low storage temperatures.
The pump requires 30 minutes to cool down to its normal operating temperature from high storage temperatures.
Do not operate until the pump is within its operating temperature range.
Ambient Air Pressure
Storage and Operating atmospheric pressure range from 70 kPA to 106 kPA
Type of Protection Against Electrical Shock
Class II, Internally-powered Equipment
Degree of Protection Against Electrical Shock
Type BF
Mode of Operation
Continuous or programmed intermittent operation
Degree of Protection Against Ingress of Fluids
Splash-proof IP24