Section X - Specifications And Symbols; Specifications - Covidien Kangaroo Joey Manual De Instrucciones

Bomba para irrigación y alimentación enteral joey con abrazadera para palo de gotero, programable
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 192
Section X — Specifications and Symbols

Specifications

TUV Listing
Type Infusion Device
Volumetric
Pumping Mechanism
Rotary Peristaltic
Pump Sets
Kangaroo™ Joey Feed-Only Set or Feed and Flush Set.
Feeding Formula Delivery Rate
1-400 mL/hr in 1 mL increments
Feeding Formula VTBD
1-3000 mL in 1 mL increments
Bolus Volume
1-3000 mL in 1 mL increments
Number of Boluses
1-99
Bolus Interval
1-24 hours in 1-hour increments
Flushing Solution Dose Range
10-500 mL in 1 mL increments
Flushing Solution Interval Range
1-24 hr in 1 hr increments
Accuracy
7% or 0.5 mL/hr, whichever is larger, with top of the fluid column at 15.24 cm (6") ± 0.76 cm (0.3") above the top of the closed
pump door, at a room temperature of 22°C ± 2°C (72°F ± 3°F), using a new Kangaroo™ Joey feeding set for no longer than the
recommended 24 hours of maximum usage. Confidence intervals for accuracy are based upon those included in the ANSI/AAMI
ID26-1992 American National Standard for Infusion Devices. Reported accuracy is based upon the time-based accuracy calculation in
this standard, as applicable for enteral feeding. Since feeding accuracy of the Kangaroo™ Joey enteral feed and flush pump with pole
clamp is based upon formula and not water, a 4% accuracy shift should be applied when testing the Kangaroo™ Joey enteral feed
and flush pump with pole clamp using water. (Water will appear to overdeliver by 4%.)
Single Fault Condition Effect on Accuracy - In the event of a single fault short circuit in the pump electronics, a maximum bolus of
1.67 mL of extra fluid delivery may occur during feeding and 8.19 ml during flushing or priming.
Kangaroo™ Joey Enteral Feed and Flush Pump with Pole Clamp
Medical Electrical Equipment
Kangaroo™ Joey Enteral Feed and Flush Pump with Pole Clamp
(1) Classified with respect to electrical shock, fire, and mechanical hazards in accordance with
ES60601-1:2012, UL60601-1, IEC 60601-1:2012.
(2) Classified with respect to electrical shock, fire, mechanical and other specified hazards in accordance
with CAN/CSA C22.2 No. 60601-1-08.
30
Table of Contents

Publicidad

Tabla de contenido
loading

Tabla de contenido