Komet E-Drive Manual Del Operador página 94

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

7 Conditioning
40
1. Brush the instrument under running water (< 38 °C, < 100 °F, at
least drinking water quality) and good lighting (min. 500 lux) and
color rendering index (min. 80 Ra) until no more dirt can be seen, for
at least 10 seconds.
2. Flush the drive channels with spray.
3. Conduct thermal disinfection or unwrapped steam sterilization.
4. Maintain mechanical parts manually [ → 41].
5. Maintain the push button chuck manually [ → 43].
6. Pack the instrument in packaging material suitable for sterilization
and storage. e.g. paper/plastic composite packaging.
7. Conduct sterilization [ → 45].
Operating Instructions E-Drive
62 61 957 D3436.201.05.05.

Publicidad

Tabla de contenido
loading

Productos relacionados para Komet E-Drive

Tabla de contenido